Abstract:AIM: To evaluate the convenience and accuracy of a novel smartphone-assisted “any-point two-step method” for finding the target axial position in cataract phacoemulsification combined with intraocular lens(IOL)implantation.
METHODS: Prospective observational study. A total of 62 cases(62 eyes)of patients with age-related cataracts who underwent cataract phacoemulsification combined with IOL implantation in our hospital from October 2021 to April 2022 were selected. They were randomly divided into two groups: 31 cases(31 eyes)in the control group were applied with the “traditional two-step method” using slit lamp to mark the target axial position of the IOL, and 31 cases(31 eyes)in the experimental group were applied with the smartphone-assisted “two-step method” to mark the target axial position of the IOL. The Callisto eye navigation system was used as a standard reference, and the deviation of the reference marking point(deviation-1), the deviation of the target axial marking point(deviation-total), and the deviation of the angle from the reference marking point to the target axial marking point(deviation-2)were calculated and recorded as the preoperative axial marking time.
RESULTS:Both deviation-1 and deviation-total values were lower in the experimental group than those in the control group(1.06°±1.39° vs 2.48°±2.23°, 1.77°±1.54° vs 2.81°±1.58°, all P<0.01), but there was no significant difference in the deviation-2 values between the two groups(1.35°±1.40° vs 1.48°±1.79°, P>0.05). The preoperative axial marking took shorter time in the experimental group than in the control group(1.77±1.70 min vs 2.88±3.20 min, P<0.01).
CONCLUSION: The smartphone-assisted “any-point two-step method” for finding the target axial position in cataract phacoemulsification combined with IOL implantation is simple, time-saving, and accurate compared with the “traditional two-step method”.