MRI in silicone oil eyes axial length measurement
Author:
Corresponding Author:

Affiliation:

Clc Number:

R770.4

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    ·AIM:To evaluate the magnetic resonance imaging(MRI)measurement accuracy of axial length in silicone oil combined cataract eye.·METHODS:A prospective double-blind controlled study,respectively with MRI and A ultrasonometry,was conducted on silicone oil combined cataract eyes to measure the axial length,and the accuracy of the two methods.·RESULTS:Thirty-two eyes achieved correct measure-ment of axial length in the plane by MRI,the axial length of MRI was longer than that of A ultrasonometry by an average length of 0.16±0.20mm.·CONCLUSION:MRI measurement of axial length is more accurate and reliable than the A ultrasonometry measurement.·