Study on the positions of different PC-IOL implantations by ultrasonic biomicroscope
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM:To evaluate the stability using ultrasonic biomicroscope(UBM) in eyes having phacoemulsification with implantation of 2 types posterior chamber intraocular lens (PC-IOL). METHODS:The degree of PC-IOL decentration,tilt and anterior chamber depth(ACD) were measured by UBM,while locating the haptics and observing the relationships of haptics on uveal tissue. RESULTS:There was no significant change during the 1 month follow-up period in decentration or tilt in each group (P>0.05).There was no significant difference among the groups in IOL decentration or tilt at any time point (P>0.05).There was no significant change during the 1 month follow-up period in ACD in each group (P>005). There was significant difference among the groups in ACD at 1 week,1 month after surgery (P<0.05). CONCLUSION:The stability of 1-piece IOL in eyes was better than that of 3-piece IOL;UBM can locate the haptics and observe the relationships of haptics on uveal tissues better.

    Reference
    Related
    Cited by
Get Citation

Shi-Yi Wang, Qian Tan. Study on the positions of different PC-IOL implantations by ultrasonic biomicroscope. Guoji Yanke Zazhi( Int Eye Sci) 2012;12(3):479-481