Comparison of central corneal thickness measured with different pachymetries
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To investigate the differences in central corneal thickness(CCT)measurements using pachymetries of ultrasound and Schempflug system.

    METHODS: The CCT of 150 healthy corneal eyes(from 86 cases)were measured with non-contact tonometry/pachymetry NT-530P, Pentacam and ultrasonic pachymetry. The consistency of the measurements was statistically analyzed by one-way ANOVA and Bland-Altman plot. The differences were analyzed by Pearson correlation coefficients.

    RESULTS: The CCT values measured with non-contact tonometry/pachymetry NT-530P, Pentacam and ultrasonic pachymetry were(541.19±32.47)μm,(542.76±32.40)μm and(539.88±31.98)μm, respectively. No statistically significance differences were found among the three instruments(F=0.290, P=0.748), and there were high correlations among them(r1=0.954, r2=0.973, r3=0.948). The sample(150 eyes)was divided into 3 groups according to the CCT measurements obtained with ultrasonic pachymetry for further analysis: CCT≤520μm,520μm 570μm. The 3 groups have high relation and good consistency(P>0.05).

    CONCLUSION: There are no statistically significant differences among the three instruments, and the correlations of them are high.

    Reference
    Related
    Cited by
Get Citation

Ya-Jing Yang, Jiang-Yue Zhao, Jin-Song Zhang. Comparison of central corneal thickness measured with different pachymetries. Guoji Yanke Zazhi( Int Eye Sci) 2012;12(10):1863-1866

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:May 10,2012
  • Revised:August 30,2012
  • Adopted:
  • Online:
  • Published: