Assessment of visual function based on IOL-Master comparing with traditional ultrasonic biometry for intraocular lens calculation in high myopia patients
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To evaluate the clinical feasibility of using a new optical coherence interferometry(IOL-Master), comparing with traditional ultrasonic biometry and manual keratometry in the accuracy and characteristics for intraocular lens calculation of high myopia.

    METHODS: The measurement of axial length was performed in 60 eyes(30 eyes for each group)with senile cataract of high myopia(≥-6.00D)using IOL-Master and ultrasonic biometry. The measurement of corneal power(K)was also performed in the patient using IOL-Master and manual keratometry preoperatively. Phacoemulsification and foldable lens implantation were done on the patients. IOL power calculation was carried out according to the SRK/T formula on the basis of the group-related data. Best corrected visual acuity, refraction, contrast sensitivity and wave front aberration root mean square(RMS)were re-tested after 3 months postoperatively.

    RESULTS: Significant difference between the two methods in axial length measurement which was 29.81±1.53mm by ultrasound and 29.63±1.81mm by IOL-Master(P=0.001). And in corneal power measurement which was 43.22±1.67K by manual keratometry and 44.27±1.39K by IOL-Master(P=0.006). There was a significant difference between the two groups(P=0.001). 63.0% vs 31.2% had a mean absolute refractive error(MARE)within ±0.50 diopter for the IOL-Master and A-scan groups, respectively(χ2=3.1, P<0.05). The RMS values of 4th order aberration, 4th order spherical aberration and total high order aberration in the IOL-Master group were lower than those in the A-scan group at 6mm pupil diameter 3 months later.

    CONCLUSION: IOL-Master is a non-contact, accurate, safe and reliable tool for calculating IOL power and it is more accurate on the design of the IOL in the cataract surgery on the high myopia patients.

    Reference
    Related
    Cited by
Get Citation

Lu Zhang. Assessment of visual function based on IOL-Master comparing with traditional ultrasonic biometry for intraocular lens calculation in high myopia patients. Guoji Yanke Zazhi( Int Eye Sci) 2013;13(10):2078-2081

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:August 16,2013
  • Revised:September 13,2013
  • Adopted:
  • Online: September 23,2013
  • Published: