Comparison of three different instruments measuring central corneal thickness of keratoconus
Author:
  • Article
  • | |
  • Metrics
  • |
  • Reference [16]
  • |
  • Related
  • |
  • Cited by
  • | |
  • Comments
    Abstract:

    AIM: To compare the difference in measurements of central corneal thickness(CCT)using A-scan,corneal specular microscopy and Pentacam in keratoconus.

    METHODS: Between July 2012 and October 2012, the CCT of 31 patients(55 eyes)were measured by A-scan, corneal specular microscopy and Pentacam. The results were analyzed by F-test and Pearson correlation.

    RESULTS: The values of CCT measured by A-scan(55 eyes), corneal specular microscopy(45 eyes)and Pentacam(52 eyes)were 469.87±57.56, 479.00±42.39, and 487.02±44.64μm, respectively; F-test results showed there were no statistical significant differences between CCT measured by three different instruments(P>0.05). The correlation between the measurements was evaluated using Pearson correlation coefficients. The CCT values by these three instruments were positively related by linear correlation analysis, A-scan and Pentacam measurements(r1=0.758, P<0.01), A-scan and Pentacam(r2=0.949, P<0.01), corneal specular microscopy and Pentacam(r3=0.685, P<0.01).

    CONCLUSION: There is a high correlation between these three instruments. A-scan and Pentacam are more precise than corneal specular microscopy. Pentacam system can measure the CCT easily, accurately and without any invasion. It is more suitable for people with keratoconus to monitor the every point of cornea and for people to do more deep research.

    Reference
    1 Tuft SJ, Moodaley LC, Gregory WM, et al. Prognostic factors for the progression of keratoconus. Ophthalmology 1994; 101(3):439-447
    2 Wollensak G. Crosslinking treatment of progressive keratoconus: new hope. Curr Opin Ophthalmol 2006; 17(4):356-360
    3 Koller T, Mrochen M, Seiler T. Complication and failure rates after corneal crosslinking. J Cataract Refract Surg 2009; 35(8):1358-1362
    4 Raiskup-Wolf F, Hoyer A, Spoerl E, et al. Collagen crosslinking with riboflavin and ultraviolet-A light inkeratoconus: Long-term results. J Cataract Refract Surg 2008; 34(5):796-801
    5 Henriquez MA, Izquierdo LJ, Bernilla C, et al. Riboflavin/Ultraviolet A corneal collagen cross-linking for the treatment of keratoconus: visual outcomes and Scheimpflug analysis.Cornea 2011; 30(3):281-286
    6张士胜,廖华萍,张静, 等. PENTACAM分析仪测量近视眼患者角膜厚度及影响因素分析.眼科新进展 2012; 32(1):44-47
    7徐艺,戴锦晖,褚仁远,等.Pentacam三维眼前节分析系统对各期圆锥角膜后表面形态的研究.眼科研究 2009; 27(3):229-233
    8 Rabinowitz YS. Keratoconus. Surv Ophthalmol 1998; 42(4):297-319
    9 Bechmann M, Thiel MJ, Neubauer AS, et al. Central corneal thickness measurement with a retinal optical coherence tomography device versus standard ultrasonic pachymetry. Cornea 2001; 20(1):50-54
    10刘祖国,张梅,陈家祺,等.圆锥角膜的角膜前后表面形态及厚度检测.中华眼科杂志 2002; 38(12):740-743
    11吴勤,段宣初,蒋幼芹,等.非接触式角膜内皮镜测量正常角膜厚度的研究.眼科学报 2004; 20(4):229-232
    12 Ho T,Cheng AC,Rao SK,et al.Central corneal thickness measurements using Orbscan II,Visante,ultrasound and Pentacam pachymetry after laser in situ keratomileusis for myopia.J Cataract Refract Surg 2007; 33(7):1177-1182
    13刘磊,栗静,王虎杰,等.三种仪器测量准分子激光原位角膜磨镶术前后中央角膜厚度的对比研究.华西医学 2012; 27(1):73-75
    14郭慧,顾宝文,杨旭,等.Pentacam和Sirius地形图系统与超声测厚仪角膜厚度的对比研究.国际眼科杂志 2012; 12(2):281-284
    15 Sanctis U, Missolungi A, Mutani B, et al. Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry. Am J Ophthalmol 2007; 144(5):712-718
    16 Grewal DS,Brar GS,Grewal SP. Assessment of central corneal thickness in normal, keratoconus, and post-laser in situ keratomileusis eyes using Scheimpflug imaging, spectral domain optical coherence tomography, and ultrasound pachymetry.J Cataract Refract Surg 2010; 36(6):954-964
    Related
    Cited by
    Comments
    Comments
    分享到微博
    Submit
Get Citation

Xu Pang, Xiu-Jun Peng, Zheng-Jun Fan. ,/et al.Comparison of three different instruments measuring central corneal thickness of keratoconus. Guoji Yanke Zazhi( Int Eye Sci) 2013;13(10):2102-2104

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:April 16,2013
  • Revised:September 11,2013
  • Online: September 23,2013