Application of high-vaccum chop combined with intermittent-energy-release technique in phacoemulsification of cataract with hard nucleus
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM:To study the effect of high-vaccum-chop combined with intermittent-energy-release technique in phacoemulsification for hard nuclear cataract.

    METHODS: Sixty-eight patients(82 eyes)with age-related cataract were included in our study, all the eyes had hard nucleus classified as Ⅳ or Ⅴ. High-vaccum-chop combined with intermittent-energy-release technique were adopted in phacoemulsification of cataract, and IOL were implanted at the same time. Time of phacoemulsification, corneal transparency after operation and visual outcome were recorded.

    RESULTS: The mean relative ultrasound power was 25%-60% intraoperative. The time of ultrasound phacoemulsification was 58-105s, mean 85±20s. On the first day after operation, corneal edema classified at grade Ⅰ were seen in 28 eyes,45 eyes with corneal edema classified at grade Ⅱ, and 9 eyes at grade Ⅲ, no eye at grade Ⅳ. Seventy-three eyes with visual acuity ≥0.3 the first day after operation, and 32 eyes with a visual acuity ≥0.5. One week after operation, best corrected visual acuity >0.5 were seen in 78 eyes, and 71 eyes with a visual acuity better than 0.8.

    CONCLUSION: High-vaccum-chop combined with intermittent-energy-release technique in ultrasonic emulsification instrument with non-burst mode is an effective and convenient way for phacoemulsification of cataract with hard nucleus, and is especially suitable for primary hospitals.

    Reference
    Related
    Cited by
Get Citation

Yue-Li Guo, Zhen-Yi Zhang, Chun-Hua Yang, et al. Application of high-vaccum chop combined with intermittent-energy-release technique in phacoemulsification of cataract with hard nucleus. Guoji Yanke Zazhi( Int Eye Sci) 2013;13(10):2107-2108

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:June 27,2013
  • Revised:September 11,2013
  • Adopted:
  • Online: September 23,2013
  • Published: