Signal filtering and parameter variance analysis from flash electroretinogram
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To accurately measure the implicit time and amplitude of a-, b-wave from flash electroretinogram(ERG)through filtering technology, eliminate oscillatory potentials interference.

    METHODS: Full-field ERGs were recorded in 30 eyes of 15 physical check-ups, measured the implicit time and amplitude of a-, b-wave, when the passband was set at 0.6~300Hz and 0.6~70Hz, and correlation was performed among those results by paired t-test.

    RESULTS: When the passband was set at 0.6~70Hz, a-, b-wave had a single peak, compared with the passband was set at 0.6~300Hz, the implicit time of a-, b-wave was prolonged, amplitude was decreased(P<0.01).

    CONCLUSION: A passband of 0.6~70Hz was the best choice to obtain smooth a- and b-waves from the original ERG, It could accurately measure the implicit time and amplitude of a-, b-wave.

    Reference
    Related
    Cited by
Get Citation

Yan-Hong Fang, Qiong-Hua Wang, Wen-Jun Gou, et al. Signal filtering and parameter variance analysis from flash electroretinogram. Guoji Yanke Zazhi( Int Eye Sci) 2014;14(11):2031-2033