Consistency analysis of Keratograph and traditional methods to evaluate tear film function
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

Ordinary Medical Scientific Research Project of Health and Family Planning Commission of Hainan Province(No.2014-055)

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To investigate repeatability and accuracy of a latest Keratograph for evaluating the tear film stability and to compare its measurements with that of traditional examination methods.

    METHODS: The results of noninvasive tear film break-up time(NI-BUT)including the first tear film break-up time(BUT-f)and the average tear film break-up time(BUT-ave)were measured by Keratograph. The repeatability of the measurements was evaluated by coefficient of variation(CV)and intraclass correlation coefficient(ICC). Wilcoxon Signed-Rank test was used to compare NI-BUT with fluorescein tear film break-up time(FBUT)to confirm the correlation between NI-BUT and FBUT, Schirmer I test values. Bland-Altman analysis was used to evaluate consistency.

    RESULTS: The study recruited 48 subjects(48 eyes)(mean age 38.7±15.2 years). The CV and ICC of BUT-f were respectively 12.6% and 0.95, those of BUT-ave were 9.8% and 0.96. The value of BUT-f was lower than that of FBUT. The difference had statistical significance(6.16±2.46s vs 7.46±1.92s, P<0.01). There was significant positive correlation between NI-BUT and FBUT, Schirmer I test values(P<0.01). The scope of 95% limits of agreement(LoA)was 4.46s in BUT-f and FBUT, while the scope of LoA was 3.64s in BUT-ave and FBUT.

    CONCLUSION: Keratograph can provide NI-BUT data that has a better repeatability and reliability, which has great application prospects in diagnosis and treatment of dry eye and refractive corneal surgery.

    Reference
    Related
    Cited by
Get Citation

Pei-Yang Shen, Hai-Bo Chen, Hong-Shan Liu, et al. Consistency analysis of Keratograph and traditional methods to evaluate tear film function. Guoji Yanke Zazhi( Int Eye Sci) 2015;15(5):846-849

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:January 22,2015
  • Revised:April 21,2015
  • Adopted:
  • Online: May 05,2015
  • Published: