Comparison of axial length and anterior chamber depth obtained by A scan ultrasonography and IOL Master in different axial length groups
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To evaluate the repeatability of axial length(AL)and anterior chamber depth(ACD)obtained by A scan ultrasonography, and to compare AL and ACD obtained by A scan with those obtained by IOL Master.

    METHODS: Two hundred and fifty-seven cataract eyes of 170 patients were included. IOL Master and A scan were performed for each eye. Five measurements of IOL Master and 3 measurements of A scan were obtained. All the tested eyes were divided into 5 groups according to AL obtained by A scan: Group A(2129mm, 21 eyes). Cronbach's Alpha coefficient and intraclass correlation coefficient(ICC)were applied to evaluate the repeatability of AL and ACD obtained by A scan. Paired t test and Pearson correlation coefficient were used to analyze the differences and correlations of AL and ACD obtained by the 2 devices, respectively. Bland-Altman plots were presented to analyze the agreements of AL and ACD obtained by the 2 devices.

    RESULTS: All the Cronbach's Alpha and ICCs of AL and ACD values were more than 0.98. The differences of AL values between A scan and IOL Master were -0.11±0.08mm in Group A, -0.15±0.10mm in Group B, -0.19±0.15 mm in Group C, -0.29±0.16mm in Group D and -0.45±0.29mm in Group E, respectively(all P<0.01). The differences of ACD values between A scan and IOL Master were -0.10±0.16mm in Group A, -0.06±0.13mm in Group B, -0.06±0.13mm in Group C, -0.19±0.10mm in Group D, -0.18±0.21mm in Group E, respectively(all P<0.01). In all groups, the AL and ACD values obtained by A scan and IOL Master presented good correlations(all r>0.89, all P<0.01).

    CONCLUSION: The AL and ACD values in cataract eyes obtained by A scan were repeatable. The AL and ACD values obtained by A scan were smaller than those obtained by IOL Master. With the increase of AL values, the differences of AL values between A scan and IOL Master increased.

    Reference
    Related
    Cited by
Get Citation

Feng-Miao Zhuang, Qiu-Yi Xiao, Yan-Jun Hua. Comparison of axial length and anterior chamber depth obtained by A scan ultrasonography and IOL Master in different axial length groups. Guoji Yanke Zazhi( Int Eye Sci) 2017;17(9):1650-1654

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:April 04,2017
  • Revised:July 31,2017
  • Adopted:
  • Online: August 22,2017
  • Published: