Comparing the corneal curvatures obtained from three different keratometers-IOL Master, Bausch & Lomb Manual keratometer and TOPCON KR-8800 autokeratometer
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To compare the corneal curvature and to investigate the agreement between three different keratometers.

    METHODS: In this prospective study, keratometry was performed using an IOL Master(IM), a Bausch & Lomb manual keratometer(Man)and TOPCON KR-8800 autokeratometer(Top)on 252 eyes of patients recruited from camps for cataract surgery. The average keratometry values were recorded and compared. The agreements between the instruments were analyzed using the Bland-Altman statistical method. The main outcome measure was average keratometry values.

    RESULTS:1)IOL Master and Manual keratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.60±1.52 D with the Manual keratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and Manual keratometer(P=0.001). The 95% LOAs of the two devices were -0.22 to 0.22 as shown in the Bland-Altman plot; 2)IOL Master and autokeratometer: the mean corneal power was 44.62±1.52 D with the IOL Master and 44.46±1.53 D with the autokeratometer. The paired t-test demonstrated a statistically significant difference in the mean corneal power between the IOL Master and autokeratometer(P<0.0001). The 95% LOAs of the two devices were -0.24 to 0.55 as shown in the Bland-Altman plot; 3)Autokeratometer and Manual keratometer: the mean corneal power was 44.60±1.52 D with Manual keratometer and 44.46±1.53 D with the autokeratometer. The paired ttest demonstrated a statistically significant difference in the mean corneal power between the autokeratometer and Manual keratometer(P<0.0001). The 95% LOAs of the two devices were -0.30 to 0.57 as shown in the Bland-Altman plot.

    CONCLUSION: Keratometry data obtained with different instruments may not be interchangeable, a fact that has important implications for cataract surgeons with respect to both surgical planning and outcomes auditing.

    Reference
    Related
    Cited by
Get Citation

Arpitha Pereira, Ajita Sasidharan. Comparing the corneal curvatures obtained from three different keratometers-IOL Master, Bausch & Lomb Manual keratometer and TOPCON KR-8800 autokeratometer. Guoji Yanke Zazhi( Int Eye Sci) 2018;18(1):17-20

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:April 01,2017
  • Revised:November 28,2017
  • Adopted:
  • Online: December 18,2017
  • Published: