Comparison of Pentacam and IOL Master for measurement of anterior segment parameters in eyes with cataracts
Author:
  • Article
  • | |
  • Metrics
  • |
  • Reference [21]
  • |
  • Related [20]
  • | | |
  • Comments
    Abstract:

    AIM: To compare anterior segment parameters in eyes with cataracts and different axial lengths measured with Pentacam and IOL Master.

    METHODS: This study included 170 eyes of 143 patients diagnosed as cataract at Tangdu Hospital, Air Force Medical University. For each patient, anterior segment parameters were measured before surgery with Pentacam and IOL Master system, respectively.

    RESULTS: There were no statistically significant differences in K1 and K2 values of eyes with long axial length(P>0.05). And there were statistically significant differences in K2 values(P<0.01), but not K1 values(P>0.05)of eyes with short and middle axial lengths. There were statistically significant differences in ACD values of eyes with any axial length(P<0.01). The Bland-Altman showed that ACD values of eyes with any axial length and K2 values of eyes with long and middle axial lengths were both in good agreement, and K1 values of eyes with any axial length were in great agreement.

    CONCLUSION: Pentacam and IOL Master can be interchanged for keratometry in eyes with long axial length, but the measurements of keratometry in eyes with short and middle axial lengths and ACD of eyes with any axial length should be compared in clinical application.

    Reference
    1 Zhang J, Zhang SS, Yu Q, et al. Comparison of anterior chamber depths measured using the Pentacam,the IOL Master,and ultrasound pachymetry. Int Eye Sci 2015; 8(15):1313-1318
    2周琳,张健. Pentacam 在白内障手术中的应用. 国际眼科杂志2015; 15(8):1358-1362
    3 Koeppl C, Findl O, Kriechbaum K, et al. Postoperative change in effective lens position of a 3-piece acrylic intraocular lens. J Cataract Refract Surg 2003; 29(10):1974-1979
    4 Hosny M, Alio JL, Claramonte P, et al. Relationship between anterior chamber depth, refractive state, corneal diameter, and axial length. J Refract Surg 2000; 16(3):336-340
    5 Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound bathymetry. J Cataract Refract Surg 2005; 31(9):1729-1735
    6 Lackner B, Schmidinger G, Skorpik C. Validity and repeat-ability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci 2005; 82(9):858-861
    7 Wegener A, Laser-Junga H. Photography of the anterior eye segment according to Scheimpflug's principle: options and limitations-a review. Clin Exp Ophthalmol 2009; 37(1):144-154
    8宋超,孟觉天. IOL Master 与 Pentacam 对角膜曲率及前房深度测量的比较. 临床眼科杂志2012; 20(1): 5-7
    9杨君,郝更生. Pentacam 与 IOLMaster 测量前房深度的一致性和可重复性研究. 临床眼科杂志2012; 20(3):208-210
    10 Nemeth G, Hassan Z, Modis L, et al. Comparision of anterior chamber depth measurement conducted with Pentacam HR© and IOLMaster©. Ophthalmic Surg Lasers Imaging 2011; 42(2):144-147
    11 Savant V, Chavan R, Pushpoth S, et al. Comparability and intra-/interobserver reliability of anterior chamber depth measurement with the Pentacam and IOLMaster. J Refract Surg 2008; 24(6): 615-618
    12 Reuland MS, Reuland AJ, Nishi Y, et al. Corneal radii and anterior chamber depth measurements using the IOL Master versus the Pentacam. J Refract Surg 2007; 23(4):368-373
    13 Dominguez-Vicent A, Perez-Vives C, Ferrer-BlascoT, et al. Interchangeability among five devices that measure anterior eye distances. Clin Exp Optom 2015; 98(5):254-262
    14潘虹,王利华,马鲁新,等. Pentacam与IOLMaster测量角膜曲率及前房深度的比较. 山东大学学报2008; 46(6):624-627
    15 Kiraly L, Stange J, Kunert KS, et al. Repeatability and agreement of central corneal thickness and keratometry measurements between four different devices. J Ophthalmol 2017; 5(3):1-8
    16宾莉,陈晓莉,李臻,等. 两种仪器测量角膜曲率及其半径的效果比较. 西部医学2017; 29(2):261-264
    17 Pelin O, Erhan O. Agreement between swept-source optical biometry and scheimpflug-based topography measurements of anterior segment parameters. Am J Ophthalmol 2016; 169(9):73-78
    18 Sel S, Stangeb J, Kaisera D, et al. Repeatability and agreement of Scheimpflug-based and swept-source optical biometry measurements. Cont Lens Anterior Eye 2017; 40(5):318-322
    19 Elbaz U, Arkana Y, Gerber Y, et al. Comparison of different techniques of anterior chamber depth and keratometric measurements. Am J Ophthalmol 2007; 143(1):48-53
    20 Dinc U, Gorgun E, Oncel B, et al. Assessment of anterior chamber depth using Visante optical coherence tomography slitlamp optical coherence tomography, IOL Master, Pentacam and OrbscanIIz. Ophthalmologica 2010; 224(6):341-346
    21 Dong J, Tang M, Zhang Y, et al. Comparison of anterior segment biometric measurements between pentacam HR and IOLMaster in normal and high myopic Eyes. PloS One 2015; 10(11):e0143110
    Cited by
    Comments
    Comments
    分享到微博
    Submit
Get Citation

Yu-Huan Yang, Hong Yan. ,/et al.Comparison of Pentacam and IOL Master for measurement of anterior segment parameters in eyes with cataracts. Guoji Yanke Zazhi( Int Eye Sci) 2019;19(5):796-800

Copy
Share
Article Metrics
  • Abstract:1189
  • PDF: 1147
  • HTML: 0
  • Cited by: 0
Publication History
  • Received:December 21,2018
  • Revised:March 29,2019
  • Online: April 22,2019