Analysis of corneal biological parameters measured by IOL Master 700 and Pentacam in patients with cataract of different ages
Author:
Fund Project:

National Key Subjects of the 13th Five-Year Plan of the National Health Commission(No.YYWS1609); Weifang Science and Technology Bureau(No.2020YX065)

  • Article
  • | |
  • Metrics
  • |
  • Reference [19]
  • |
  • Related [20]
  • | | |
  • Comments
    Abstract:

    AIM:To compare the difference, correlation and consistency of corneal biological parameters measured by IOL Master 700 and Pentacam before cataract extraction and intraocular lens implantation in different age groups.

    METHODS: Cross-sectional study. A total of 87 cataract patients(162 eyes)in Weifang Eye Hospital from February to September 2020 were collected, including 44 males(80 eyes)and 43 females(82 eyes)(age 61.2±9.87 years). The patients were divided into four groups: group A \〖17 cases(32 eyes), 40-50 years old\〗, group B \〖25 cases(47 eyes), 51-60 years old\〗, group C \〖28 cases(53 eyes), 61-70 years old\〗, and group D \〖17 cases(30 eyes), 71-80 years old\〗. The preoperative corneal biological parameters of cataract patients were measured by IOL Master 700 and Pentacam, and the flat axis corneal curvature(K1), steep axis corneal curvature(K2), mean corneal curvature(Km), corneal astigmatism, anterior chamber depth(ACD)and central corneal thickness(CCT)were recorded. The difference and correlation of measurement results between two kinds of biometric instruments in different age groups were analyzed.

    RESULTS: Except for the corneal astigmatism in group C and K1 and corneal astigmatism in group D, there were significant differences between the two instruments(t=2.746, -2.582, 2.637, all P<0.05), but there was no significant difference in other parameters among the four groups. Pearson correlation analysis showed that there was a good correlation between the two instruments in measuring the parameters of the four groups of patients. The results of Bland-Altman analysis showed that there was a good consistency between IOL Master 700 and Pentacam in measuring K1, K2, Km, corneal astigmatism, ACD and CCT in the four groups.

    CONCLUSION:There was no significant difference between IOL Master 700 and Pentacam in the measurement of corneal biological parameters in cataract patients aged 40-60 years, but there was significant difference in astigmatism between 61-70 years old, astigmatism and K1 value in 71-80 years old patients. Pearson correlation analysis showed that there was a good correlation between the two instruments. Generally speaking, the consistency of the two kinds of examination equipment is good, and the corneal astigmatism and corneal curvature should be selected by comprehensive analysis of the data.

    Reference
    1 Lundström M, Pesudovs K. Questionnaires for measuring cataract surgery outcomes. J Cataract Refract Surg 2011; 37(5):945-959
    2 Lee AC, Qazi MA, Pepose JS. Biometry and intraocular lens power calculation. Curr Opin Ophthalmol 2008; 19(1):13-17
    3 Tuncer I, Zengin MÖ, Yıldız S. The effect of cycloplegia on the ocular biometry and intraocular lens power based on age. Eye (Lond)2021; 35(2):676-681
    4 Petermeier K, Gekeler F, Messias A, et al. Intraocular lens power calculation and optimized constants for highly myopic eyes. J Cataract Refract Surg 2009; 35(9):1575-1581
    5 Kaswin G, Rousseau A, Mgarrech M, et al. Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard. J Cataract Refract Surg 2014; 40(4):593-600
    6 Savini G, Taroni L, Schiano-Lomoriello D, et al. Repeatability of total Keratometry and standard Keratometry by the IOL Master 700 and comparison to total corneal astigmatism by Scheimpflug imaging. Eye (Lond)2021; 35(1):307-315
    7 Kunert KS, Peter M, Blum M, et al. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg 2016; 42(1):76-83
    8刘勇, 扬楚春, 汪键, 等. 不同年龄组的角膜厚度及角膜曲率的变化规律. 国际眼科杂志 2009; 9(1):73-74
    9曹建雄, 刘伟, 刘妍, 等. Pentacam与IOL master测量角膜屈光力的比较. 国际眼科杂志 2019; 19(1):113-117
    10 Hashemi H, Asgari S, Emamian MH, et al. Age-related changes in corneal curvature and shape: the shahroud eye cohort study. Cornea 2015; 34(11):1456-1458
    11 Eibschitz-Tsimhoni M, Tsimhoni O, Archer SM, et al. Effect of axial length and keratometry measurement error on intraocular lens implant power prediction formulas in pediatric patients. J AAPOS 2008; 12(2):173-176
    12杨玉焕, 严宏. Pentacam与IOL master对白内障眼前节参数测量的比较. 国际眼科杂志 2019; 19(5):796-800
    13 Anderson DF, Dhariwal M, Bouchet C, et al. Global prevalence and economic and humanistic burden of astigmatism in cataract patients: a systematic literature review. Clin Ophthalmol Auckl N Z 2018; 12:439-452
    14 Savant V, Chavan R, Pushpoth S, et al. Comparability and intra-/interobserver reliability of anterior chamber depth measurements with the Pentacam and IOL Master. J Refract Surg 2008; 24(6):615-618
    15吴怡, 唐少华, 杨光, 等. 三种方法测量白内障患者前房深度的比较及其影响因素. 国际眼科杂志 2021; 21(4):702-706
    16 Muzyka-Woz'niak M, Oleszko A. Comparison of anterior segment parameters and axial length measurements performed on a Scheimpflug device with biometry function and a reference optical biometer. Int Ophthalmol 2019; 39(5):1115-1122
    17 Elbaz U, Barkana Y, Gerber Y, et al. Comparison of different techniques of anterior chamber depth and keratometric measurements. Am J Ophthalmol 2007; 143(1):48-53
    18 Dinc UA, Gorgun E, Oncel B, et al. Assessment of anterior chamber depth using Visante optical coherence tomography, slitlamp optical coherence tomography, IOL Master, Pentacam and Orbscan IIz. Ophthalmologica 2010; 224(6):341-346
    19 Hashmani N, Hashmani S, Hanfi A, et al. Effect of age, sex, and refractive errors on central corneal thickness measured by Oculus Pentacam©. Clin Ophthalmol 2017; 11:1233-1238
    Cited by
    Comments
    Comments
    分享到微博
    Submit
Get Citation

Shan Ma, Yan-Chun Ma, Xiu-Hua Liu,/et al.Analysis of corneal biological parameters measured by IOL Master 700 and Pentacam in patients with cataract of different ages. Guoji Yanke Zazhi( Int Eye Sci) 2021;21(10):1792-1797

Copy
Share
Article Metrics
  • Abstract:613
  • PDF: 1092
  • HTML: 0
  • Cited by: 0
Publication History
  • Received:June 08,2021
  • Revised:September 08,2021
  • Online: September 16,2021