    Abstract:

    AIM: To investigate the safety and efficacy of Paul glaucoma implant(PGI)in the short-term follow-up period and share first experience with this novel aqueous shunt in Indonesian populations.

    METHODS: A total of 21 patients(22 eyes)with PGI implants from April 2022 to December 2022 and with at least a complete 2mo follow-up were retrospectively analyzed. The primary outcome measure was failure, defined as intraocular pressure(IOP)out of the target range of 21 mmHg or less than 20% reduction from baseline for 2 consecutive visits, other glaucoma surgeries required, or removal of the implant.

    RESULTS: The follow-up period was 2 to 6mo. The mean IOP reduction was 52.27±22.94%, with a range of 9% to 90%. The complete success rate was 59%, and patients with or without a history of glaucoma surgery had 50% and 59% of complete success rates, respectively. Complications of the surgery were diplopia(n=2), early hypotony(n=1), hyphema(n=1), and exposed tube(n=2).

    CONCLUSION: The complete success of the PGI implantation was 57%. No serious postoperative complications were found in our cases. One case of hypotony resolved in the early postoperative period.

    Reference
    Related
    Cited by
Get Citation

Emma Rusmayani, Viona Viona, Iwan Soebijantoro, et al. Early experience with the novel glaucoma shunt device: Paul glaucoma implant in the Indonesian populations. Guoji Yanke Zazhi( Int Eye Sci) 2023;23(10):1603-1608

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:June 15,2023
  • Revised:August 04,2023
  • Adopted:
  • Online: September 19,2023
  • Published: