Agreement of corneal high-order aberrations measured by three wavefront aberrometers in myopic adults
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

Guangxi Science and Technology Project(No.GuiKeAD19245193)

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM:To evaluate the agreement of corneal high-order aberrations from Topcon KR-1W, i.Profiler and OPD-Scan Ⅲ wavefront aberrometers in myopic adults.

    METHODS:A prospective clinical study. A total of 92 adult patients(92 eyes)with myopia in the department of optometry, the People's Hospital of Guangxi Zhuang Autonomous Region from June to August 2022 were enrolled. The third-order and fourth-order corneal aberrations at the pupil diameter of 4 and 6mm were measured by Topcon KR-1W, i.Profiler, and OPD-Scan Ⅲ, respectively. The difference and agreement of the three aberrometers were evaluated.

    RESULTS: The measurements at 6mm pupil diameter were all greater than those at 4mm pupil diameter. Although there were no statistical differences in the measurements of Z-44、Z-24 by the three aberrometers at 4 pupil diameter(P>0.05), there were statistical differences in other measurements(P<0.05). The aberration results measured by the three aberrometers were statistically different at the 6mm pupil diameter(P<0.05). The 95% limit of agreement(95%LoA)of the measurements of higher-order aberration, including the third-order aberrations at 4mm pupil diameter and the third-order and fourth-order aberrations at 6mm pupil diameter(except for the Z-24)were greater than 0.1μm. The concordance correlation coefficient(Pc)was lower than 0.90, indicating a poor consistency. The correlation coefficients of corneal higher-order aberrations were significantly different among the three aberrometers at 4 and 6mm pupil diameter(r4mm=0.215~0.805, P4mm<0.05; r6mm=0.561~0.916, P6mm<0.001).

    CONCLUSION:There were significant differences in the measurements of the third- and fourth-order corneal aberrations at 4 and 6mm pupil diameter among Topcon KR-1W, i.Profiler, and OPD-Scan Ⅲ, and the agreements were poor, so they are not interchangeably in clinical applications.

    Reference
    Related
    Cited by
Get Citation

Hua-Xian Zou, Qi Chen, Li-Li Li, et al. Agreement of corneal high-order aberrations measured by three wavefront aberrometers in myopic adults. Guoji Yanke Zazhi( Int Eye Sci) 2023;23(10):1723-1731

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:June 06,2023
  • Revised:August 24,2023
  • Adopted:
  • Online: September 19,2023
  • Published: