Accuracy assessment of refractive status in patients implanted with extended depth of focus intraocular lens
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

Clinical Research and Cultivation Project of Shanghai Eye Diseases Prevention & Treatment Center(No.21LC01003, 21LC01007)

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To evaluate the refractive status through computer refractometer and OPD-Scan III auto refractometer in cataract patients after extended depth of focus(EDOF)intraocular lens implantation.

    METHODS: Retrospective observational study. A total of 61 cases(76 eyes)that received phacomulsification and implanted with TECNIS® Symfony ZXR00 intraocular lens in Shanghai Eye Diseases Prevention & Treatment Center from May 2022 to May 2023 were collected. Measurements from the computer refractometer, OPD-Scan III auto refractometer, and subjective refraction, were taken from all patients on the same day postoperatively.

    RESULTS: There were statistical significant difference in sphere(S)and spherical equivalent(SE)readings from the computer refractometer and subjective refraction(all P<0.01), with mean differences of -0.67±0.37 D and -0.75±0.35 D, respectively, and the S and SE obtained from computer refractometer more incline to myopia than those from subjective refraction; there were statistical significant difference in computer refractometer and subjective refraction(P<0.01), with a relative small absolute difference(0.21±0.24 D). The S, cylinder(C)and SE of computer refractometer(S, C, SE)were positively correlated with subjective refraction(r=0.7994, 0.7929, and 0.8118, respectively, all P<0.01). Additionally, there were statistical significant differences in S, C and SE of OPD-Scan Ⅲ and subjective refraction(P<0.01), and the absolute differences of S(0.63±0.36 D), C(0.35±0.26 D)and SE(0.53±0.36 D)were small. Furthermore, the S, C and SE of OPD-Scan Ⅲ were positively correlated with subjective refraction(r=0.4410, 0.4982, 0.5224, all P<0.01).

    CONCLUSION: In patients who received implantation of EDOF lenses, the consistency of computer refractometer, OPD-Scan III auto refractometer and subjective refraction was good. The average difference of the S and SE obtained via computer refractometer was large, but both exhibited a myopic shift relative to those derived from subjective refraction, and the C values demonstrated minimal discrepancy. Furthermore, the differences between OPD-Scan III auto refractometer and subjective refraction were small, but the direction of the difference is unstable, sometimes it is myopic deviation, while sometimes it is hyperopic deviation.

    Reference
    Related
    Cited by
Get Citation

Yu Chunxia, Fang Xiaoling, Xue Wenwen, et al. Accuracy assessment of refractive status in patients implanted with extended depth of focus intraocular lens. Guoji Yanke Zazhi( Int Eye Sci) 2024;24(11):1821-1825

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:February 22,2024
  • Revised:September 27,2024
  • Adopted:
  • Online: October 18,2024
  • Published: