Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To analyze the correlation of Goldmann applanation tonometer (GAT), I-Care tonometer and Tono-Pen tonometer results in young healthy persons, and to investigate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements recorded with these tonometers. METHODS: We conducted a pilot clinical study in 78 eyes of 78 subjects aged 22-28 years old (44 women and 34 men; mean age 23.8±1.19y). IOP was measured using GAT, I-Care and Tono-Pen tonometers, followed by measurements of CCT. Statistical analysis was performed using SPSS 20.0. RESULTS: The mean IOPs and standard deviation (±SD) for GAT, I-Care and Tono-Pen were 15.62±2.281 mm Hg, 16.29±2.726 mm Hg and 16.32±2.393 mm Hg, respectively. The mean CCT was 555.15±29.648 μm. Clear positive correlations between GAT and I-Care, GAT and Tono-Pen, and I-Care and Tono-Pen tonometers were found (r=0.867, P<0.001; r=0.861, P<0.001; r=0.915, P<0.001, respectively). In comparison between devices, Bland–Altman analysis showed a significant mean difference (MD) in the measurements by GAT and I-Care of ?0.679 mm Hg and by GAT and Tono-Pen of ?0.705 mm Hg ( P<0.001), but there was no significant difference between I-Care and Tono-Pen ( P>0.05). Both non-gold standard tonometers were affected by CCT; that is, both I-Care and Tono-Pen tonometer values were significantly higher with higher CCT means (>555 μm; MD=?1.282, P<0.001; MD=?0.949, P<0.001, respectively) compared with GAT. CONCLUSION: Both I-Care and Tono-Pen tonometers overestimated IOP compared with the GAT values. Either the I-Care or Tono-Pen tonometer could be used instead of GAT because there was no significant difference between their results. Higher CCT values (>555 μm) were associated with overestimated IOP values.

    Reference
    Related
    Cited by
Get Citation

Saulius Galgauskas, Rasa Strupaite, Ernesta Strelkauskaite, et al. Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons. Int J Ophthalmol, 2016,9(1):76-80

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:February 26,2015
  • Revised:April 29,2015
  • Adopted:
  • Online: December 29,2015
  • Published: