Citation:Jin GM,Xiao B,Zhou YJ,Wang YY,Li XP,Zheng DY.Agreement of corneal curvature and central corneal thickness obtained from a swept-source OCT and Pentacam in ectopia lentis patients.Int J Ophthalmol 2020;13(8):1244-1249,doi:10.18240/ijo.2020.08.10
Agreement of corneal curvature and central corneal thickness obtained from a swept-source OCT and Pentacam in ectopia lentis patients
Received:February 11, 2020  Revised:June 04, 2020
Email this Article  Add to Favorites  Print
DOI:10.18240/ijo.2020.08.10
Key Words:corneal curvature  Pentacam  swept-source Fourier-domain anterior segment optical coherence tomography  ectopia lentis
Fund Project:Supported by National Natural Science Foundation of China (No.81873673; No.81900841); the Fundamental Research Funds of the State Key Laboratory of Ophthalmology (No.30306020240020212); the Young Teachers Training Program of Sun Yat-sen University (No.20ykpy143).
                 
AuthorInstitution
Guang-Ming Jin State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou , Guangdong Province, China
Bing Xiao State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou , Guangdong Province, China
Yi-Jing Zhou State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou , Guangdong Province, China
Yi-Yao Wang State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou , Guangdong Province, China
Xue-Pei Li State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou , Guangdong Province, China
Dan-Ying Zheng State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou , Guangdong Province, China
Hits: 781
Download times: 139
Abstract:
      AIM: To assess the inter-device consistency of corneal curvature and central corneal thickness between Pentacam and a swept-source Fourier-domain anterior segment optical coherence tomography (AS-OCT) in ectopia lentis patients.

    METHODS: Totally 72 eyes of ectopia lentis patients were recruited. Central corneal thickness (CCT), corneal curvature values and corneal astigmatism were obtained from both the Pentacam and AS-OCT (CASIA2). Repeatability was evaluated for both devices. The coefficient of repeatability (COR) and the relative COR was calculated. Bland-Altman plots were conducted to evaluate the inter-device agreement of measurement. Orthogonal linear regression was used to examine any proportional bias.

    RESULTS: The mean difference of CCT, steep anterior corneal curvature (anterior KS), flat anterior corneal curvature (anterior Kf), anterior corneal astigmatism (ACA), steep posterior corneal curvature (posterior KS), flat posterior corneal curvature (posterior Kf), posterior corneal astigmatism (PCA), steep true net power (TNP KS), flat true net power (TNP Kf) and total corneal astigmatism (TCA) between Pentacam and CASIA2 were 7.03±9.70 μm, -0.19±0.41 D, -0.27±0.35 D, 0.04±0.47 D, -0.17±0.23 D, -0.11±0.11 D, -0.02±1.02 D -0.41±0.43 D, -0.52±0.46 D, and -0.15±0.96 D, respectively. For measurement of TNP Kf with the Pentacam and CASIA2, a mean difference of 0.52 D and COR of 0.90 with P=0.02 was detected. There was no significant difference in CCT (P=0.393), anterior Kf (P=0.107), anterior Ks (P=0.414), ACA (P=0.131), posterior Kf (P=0.286), posterior Ks (P=0.418), PCA (P=0.105), TNP Ks (P=0.054), and TCA (P=0.977) between Pentacam and CASIA2.

    CONCLUSION: Our study reveals good agreement of CCT, corneal curvature and corneal astigmatism measured by CASIA2 and Pentacam in ectopia lentis patients. However, there was significant difference for CCT and corneal curvature values obtained by the two devices.