飞秒激光制瓣LASIK手术后角膜生物力学变化研究
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

西安市社会发展引导项目 \〖No.SF1513(2)\〗; 陕西省社会发展项目(No.2016SF-274); 陕西卫生科研项目(No.2016E002)


Corneal biomechanical properties after femtosecond laser assisted LASIK with the corneal visualization Scheimpflug technology and ocular response analyzer
Author:
Affiliation:

Fund Project:

Supported by Xi'an Social Development Research Project Grant \〖No.SF1513(2)\〗; Shaanxi Social Development Research Project Grant(No.2016SF-274); Shaanxi Provincial Health Research Project(No.2016E002)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:应用可视化角膜生物力学分析仪(Corvis ST)及眼反应分析仪(ORA)评估飞秒激光制瓣准分子激光原位角膜磨镶术(laser in situ keratomileusis,LASIK)术后角膜生物力学变化,并分析与其他参数的相关性。

    方法:收集63例飞秒制瓣LASIK手术患者,所有收集对象采用右眼数据进行分析。角膜生物力学测量仪器应用可视化角膜生物力学分析仪Corvis ST及眼反应分析仪ORA,手术前及手术后1mo进行生物力学数据采集分析。配对t检验或Mann-Whitney U检验进行手术前后生物力学对比分析。Pearson 或Spearman统计学方法进行相关性分析。

    结果:与FS-LASIK手术前相比,手术后1st A-time,Vin,2nd A length,Vout以及 Radius 下降,有统计学差异(P=0.00, P=0.00, P=0.00, P=0.00, P=0.00)。2nd A-time,DA以及PD手术后增加,有统计学差异(P=0.00, P=0.00, P=0.00)。1st A length以及HC time 手术后改变不明显,无统计学差异(P=0.96,P=0.08)。与FS-LASIK手术前相比,手术后CH,CRF下降,有统计学差异(P=0.00, P=0.00)。手术后1st A-time, 2nd A-time, DA and Radius变化量与手术前角膜中央厚度有明显相关性(P=0.01, P=0.04, P=0.03, P=0.01)。

    结论:飞秒激光LASIK手术后角膜生物力学参数有较明显改变,可通过Corvis ST 及ORA生物力学参数计算得出相应变化,手术后生物力学参数变化与角膜厚度相关性明显。

    Abstract:

    AIM: To investigate the changes of corneal biomechanical properties before and after femtosecond laser assisted LASIK(FS-LASIK)using Corneal Visualisation Scheimpflug Technology(Corvis ST)and Ocular Response Analyzer(ORA), and the correlation with other myopic parameters.

    METHODS:Sixty three patients(63 eyes)who had myopic femtosecond laser assisted LASIK(FS-LASIK)were enrolled in the study. The right eye of each patient was analyzed in this study. The corneal biomechanical parameters pre-operative and 1mo post-operative was measured with the Corvis ST(Oculus, Wetzlar, Germany)and ORA(Reichert, Buffalo, New York, USA). Comparison of the biomechanical property values before and after surgery was peformed using Paired t-test or Mann-Whitney U. Pearson or Spearman correlations were used to evaluate the relationship between parameters.

    RESULTS: The postoperative 1st A-time, Vin, 2nd A length, Vout, HC time and Radius demonstrate significant decreases comparing with preoperative values(P=0.00, P=0.00, P=0.00, P=0.00, P=0.00, P=0.00 respectively). The postoperative 2nd A-time, DA and PD significantly increases(P=0.00, P=0.00, P=0.00), however, the 1st A length had no significant difference after surgery. The CH and CRF were significantly lower after FS-LASIK(P=0.00, P=0.00). A statistically significant correlation coefficient was found between preoperative central corneal thickness(CCT)with postoperative-preoperative changes of 1st A-time, 2nd A-time, DA and Radius respectively(P=0.01, P=0.04, P=0.03, P=0.01).

    CONCLUSION:There were significantly changes of corneal biomechanical properties after FS-LASIK surgery. The changes of corneal biomechanical properties after FS-LASIK can be reflected by some parameters of Corvis ST and ORA. The mainly influence of corneal biomechanical alteration was possibly correlation with corneal thickness.

    参考文献
    相似文献
    引证文献
引用本文

李晶,魏升升,万雅群,等.飞秒激光制瓣LASIK手术后角膜生物力学变化研究.国际眼科杂志, 2017,17(2):195-199.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2015-11-17
  • 最后修改日期:2016-12-08
  • 录用日期:
  • 在线发布日期: 2017-01-20
  • 出版日期:
文章二维码