Diaton眼压计测量准分子激光原位角膜磨镶术后眼压
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

R779.63

基金项目:


Diaton tonometer for intraocular pressure measurements after laser in situ keratomileusis
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:通过与非接触式眼压计(non-contact tonometer,NCT)的比较,评价Diaton眼压计(DT)在准分子激光原位角膜磨镶术(laser in situkeratomileusis,LASIK)前后眼压测量中的应用价值。方法:在LASIK术前及术后1mo,分别使用NCT和DT测量患者眼压,使用SPSS13.0统计软件对所得结果进行统计学分析。结果:术后1mo时,NCT眼压测量值较术前下降(5.76±2.31)mmHg,DT眼压测量值则无显著差异(t=0.70,P=0.487);NCT与DT读数测量值差值与术中切削量呈正相关,与术后中央角膜厚度呈负相关。结论:Diaton测量眼压不受角膜厚度的影响,因此更适用于LASIK术后患者的眼压测量。

    Abstract:

    AIM: To evaluate the performance of Diaton tonometer(DT) before and after laser in situ keratomileusis(LASIK) by comparing it with non-contact tonometer (NCT). METHODS:NCT and DT were used to measure the intraocular pressure before and 1 month after LASIK, the results were analyzed with SPSS 13.0. RESULTS: One month postoperatively, NCT readings declined by 5.76±2.31mmHg, DT readings did not differ significantly(t=0.70,P=0.487); the difference between NCT and DT readings had positive correlation with declined corneal thickness, and had negative correlation with the central corneal thickness postoperatively. CONCLUSION: DT readings were not affected by corneal thickness; therefore, Diaton tonometer may be a better choice for the measurement of intraocular pressure after LASIK.

    参考文献
    相似文献
    引证文献
引用本文

刘磊,栗静,李新宇,等. Diaton眼压计测量准分子激光原位角膜磨镶术后眼压.国际眼科杂志, 2010,10(4):687-688.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:
  • 最后修改日期:
  • 录用日期:
  • 在线发布日期:
  • 出版日期: