Pentacam与IOL Master对白内障眼前节参数测量的比较
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:


Comparison of Pentacam and IOL Master for measurement of anterior segment parameters in eyes with cataracts
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:比较Pentacam与IOL Master对不同眼轴眼白内障眼前节参数的测量。

    方法:选择空军军医大学唐都医院拟行白内障手术患者143例170眼,术前分别使用Pentacam和IOL Master测量眼前节参数,对两种测量结果进行比较分析。

    结果:两者对长眼轴眼K1、K2值均无差异(P>0.05),对短眼轴眼、中眼轴眼K1值均无差异(P>0.05),但对K2值有差异(P<0.01),对短、中、长眼轴的眼中央前房深度值均有差异(P<0.01)。两种测量仪器对各眼轴眼前房深度测量一致性较好,对中、长眼轴眼K2值的测量一致性也较好,对各眼轴眼的K1值测量一致性更好。

    结论:Pentacam与IOL Master对白内障术前长眼轴眼的角膜曲率测量值可以替代,对短、中眼轴眼角膜曲率和各眼轴眼的中央前房深度值的测量结果需要结合临床比较应用。

    Abstract:

    AIM: To compare anterior segment parameters in eyes with cataracts and different axial lengths measured with Pentacam and IOL Master.

    METHODS: This study included 170 eyes of 143 patients diagnosed as cataract at Tangdu Hospital, Air Force Medical University. For each patient, anterior segment parameters were measured before surgery with Pentacam and IOL Master system, respectively.

    RESULTS: There were no statistically significant differences in K1 and K2 values of eyes with long axial length(P>0.05). And there were statistically significant differences in K2 values(P<0.01), but not K1 values(P>0.05)of eyes with short and middle axial lengths. There were statistically significant differences in ACD values of eyes with any axial length(P<0.01). The Bland-Altman showed that ACD values of eyes with any axial length and K2 values of eyes with long and middle axial lengths were both in good agreement, and K1 values of eyes with any axial length were in great agreement.

    CONCLUSION: Pentacam and IOL Master can be interchanged for keratometry in eyes with long axial length, but the measurements of keratometry in eyes with short and middle axial lengths and ACD of eyes with any axial length should be compared in clinical application.

    参考文献
    相似文献
    引证文献
引用本文

杨玉焕,严宏. Pentacam与IOL Master对白内障眼前节参数测量的比较.国际眼科杂志, 2019,19(5):796-800.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2018-12-21
  • 最后修改日期:2019-03-29
  • 录用日期:
  • 在线发布日期: 2019-04-22
  • 出版日期: