连续视程人工晶状体和区域折射型人工晶状体植入术后视觉质量的对比分析
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

吉林省科技厅国际合作项目(No.20200801026GH); 吉林省卫生与健康技术创作新项目(No.2020J038)


Comparative analysis of visual quality after implantation of Symfony IOL and MF15 IOL
Author:
Affiliation:

Fund Project:

Jilin Province Science and Technology Development Program(No.20200801026GH); Jilin Health Technology Innovation Project(No.2020J038)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:比较连续视程(Tecnis Symfony)人工晶状体(IOL)和区域折射型(Lentis Comfort LS-313 MF15)IOL植入术后患者的视觉质量。

    方法:临床前瞻性研究。收集2021-01/12在我院行白内障超声乳化联合IOL植入术的白内障患者267例(305眼),按植入IOL类型分为Symfony组(A组,160眼)和MF15组(B组,145眼),观察两组患者术后3mo裸眼远视力(uncorrected distance visual acuity, UDVA)(5m)、裸眼中视力(uncorrected intermediate visual acuity, UIVA)(80cm)、裸眼近视力(uncorrected near visual acuity, UNVA)(40cm),离焦曲线,调制传递函数(MTF)和高阶相差(HOAs),包括彗差、球差、三叶草差,进行视觉质量问卷调查(Quality of Vision questionnaire,QoV)。

    结果:两组患者术前各参数差异均无统计学意义(P>0.05)。A组和B组UDVA和UNVA差异无统计学意义(P>0.05); A组的UIVA高于B组(P<0.05); A组的MTF值在各个空间频率下均高于B组,差异具有统计学意义(P<0.05); A组的HOAs低于B组(P<0.05); 术后3mo绘制离焦曲线显示0D处两组视力相似,-0.5D~-2.5D处A组视力优于B组,且在-1.0D和-2.0D处差异有统计学意义(P<0.05); A组的Qov分数高于B组(P<0.05); A组出现光晕,星爆等不良视觉干扰现象的频率要高于B组。

    结论:MF15和Symfony均提供了稳定的远,近距离视力; 与MF15相比,Symfony IOL具有较好的中距离视力,较高的对比敏感度以及较低的HOAs; 与Symfony相比,MF15人工晶状具有更少的术后不良视觉干扰现象(光晕,星芒等)。

    Abstract:

    AIM: To compare the visual quality in patients after implantation of extended-range-of-vision(Tecnis Symfony)intraocular lens(IOL)and zonal refractive(Lentis Comfort LS-313 MF15)IOL.

    METHODS: A prospective clinical study. Collecting 267 cataract patients(305 eyes)who underwent phacoemulsification combined with IOL implantation surgery in our hospital from January 2021 to December 2021,they were divided into the Symfony group(group A, 160 eyes)and MF15 group(group B, 145 eyes)according to the types of implanted IOL. Postoperative visual acuity were observed at 3mo and included uncorrected distance visual acuity(UDVA, 5m), uncorrected intermediate visual acuity(UIVA, 80 cm), uncorrected near visual acuity(UNVA, 40cm), defocus curve, modulation transfer function(MTF)and high-order aberrations(HOAs), including coma, spherical aberration and trefoil aberration, and conducted the Quality of Vision(QoV)questionnaire.

    RESULTS: There were no significant differences in the preoperative parameters in both groups(P>0.05). There was no significant difference between the UDVA and UNVA in groups A and B(P>0.05); The UIVA was higher in group A than in group B(P<0.05); The MTF value in group A was higher than group B at all spatial frequency, and the difference was statistical significant(P<0.05); The HOAs in group A was lower than that in group B(P<0.05); Defocus curves were plotted 3mo after the operation suggest that the visual acuity of the two groups was similar at 0D, and the visual acuity of the groups A was better than group B from -0.5D to -2.5D, and there was a statistically significant difference at -1.0D and -2.0D(P<0.05). The QoV score in group A was significantly higher than that in group B(P<0.05). The frequency of bad visual interference such as halo and starburst was higher in group A than in group B.

    CONCLUSION: Both the MF15 and Symfony can provide stable distance and near visual acuity. Compared with MF15, Symfony had better intermediate visual acuity, higher contrast sensitivity and lower HOAs. Compared with Symfony, MF15 IOL had less postoperative bad visual interference symptoms(halo, starburst, etc).

    参考文献
    相似文献
    引证文献
引用本文

郭榕,李卓亚,胡晓敏,等.连续视程人工晶状体和区域折射型人工晶状体植入术后视觉质量的对比分析.国际眼科杂志, 2022,22(8):1239-1244.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2022-02-13
  • 最后修改日期:2022-06-24
  • 录用日期:
  • 在线发布日期: 2022-07-27
  • 出版日期:
文章二维码