两种基于Scheimpflug成像原理的生物测量仪对眼前节参数测量的一致性研究
CSTR:
作者:
作者单位:

作者简介:

通讯作者:

中图分类号:

基金项目:

湖南省企业联合基金项目(No.2023JJ70040); 爱尔眼科集团中青年科技创新孵化培育专项计划项目(No.AC2214D01); 2021年爱尔眼科集团科研基金项目自由探索计划(No.AF2102D6); 爱尔眼科集团临床研究所科研基金项目(No.AR2102D1)


Agreement study of anterior segment parameters measured by two kinds of ocular biometry based on imaging principle of Scheimpflug
Author:
Affiliation:

Fund Project:

Enterprise Joint Fund Project of Hunan Province(No.2023JJ70040); Young and Middle-aged Science and Technology Innovation Incubation Project of Aier Eye Group(No.AC2214D01); Free Exploration Program of the Scientific Research Foundation of Aier Eye Group in 2021(No.AF2102D6); Clinical Research Institute Research Fund Project of Aier Eye Group(No.AR2102D1)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:评估两种基于Scheimpflug成像技术原理的国产Scansys和进口Sirius三维眼前节分析系统,对近视患者眼前节生物参数测量的差异性与一致性。

    方法:系列病例研究,选取2022-05/10沈阳爱尔卓越眼科医院诊断为轻中度近视拟行屈光手术(包括全飞秒、半飞秒、TransPRK、ICL植入术)患者103例103眼(所有患者均右眼入组)。收集患者术前角膜曲率(Km)、中央角膜厚度(CCT)、前房深度(ACDEndo.)、前房角(ACA)、前房容积(ACV)、白到白(WTW)。

    结果:Scansys和Sirius测量术前Km为42.88(41.54,44.60)、42.98(41.56,44.52)D,CCT为541.52±29.08、549.55±29.62μm,ACA为42.70°±2.67°、46.63°±5.13°,WTW为12.10±0.60、11.98±0.47mm,差异均有统计学意义(均P<0.01)。Scansys和Sirius测量ACV为194.26±31.06、191.47±25.65mm3,ACDEndo.为3.40(3.17,3.57)、3.43(3.19,3.56)mm,差异均无统计学意义(均P>0.05)。两种仪器测量Km、CCT、ACA、ACDEndo.、ACV、WTW值区间范围小,平均差异接近0,95%LoA线外点比例均<5%,一致性较好。

    结论:Scansys和Sirius对眼前节参数测量结果差异较小,一致性较好,临床上可以相互参考替代。Scansys理论上可以用来推算ICL型号,或可成为眼前节参数测量的一种新选择。

    Abstract:

    AIM: To compare the differences and agreement of anterior segment biometric parameters of myopic patients measured by domestic Scansys and the imported Sirius based on the principle of Scheimpflug imaging technique.

    METHODS: In this case series study, 103 cases(103 eyes)that underwent pre-refractive surgery(including small incision lenticule extraction, femtosecond laser-assisted in situ keratomileusis, transepithelial photorefractive keratectomy and implantable contact lens implantation)at Aier Excellent Eye Hospital from May 2022 to October 2022 were recruited. Preoperative keratometry(Km), central corneal thickness(CCT), anterior chamber depth(ACDEndo.), anterior chamber angle(ACA), anterior chamber volume(ACV), white to white(WTW)of patients were recorded.

    RESULTS: The results of Km, CCT, ACA, and WTW measured by Scansys and Sirius were 42.88(41.54, 44.60)and 42.98(41.56, 44.52)D,(541.52±29.08)and(549.55±29.62)μm, 42.70°±2.67° and 46.63°±5.13°, 12.10±0.60 and 11.98±0.47 mm, respectively, showing the difference was statistically significant(all P<0.01). The ACV measured by Scansys and Sirius was 194.26±31.06 and 191.47±25.65 mm3, and ACDEndo. was 3.40(3.17, 3.57)and 3.43(3.19, 3.56)mm, with no statistically significant difference(all P>0.05). The range of Km, CCT, ACA, ACDEndo., ACV and WTW values measured by the two instruments was small, with an average difference close to zero, and the points percentage of 95% limits of agreement(LoA)was <5%, which is of good consistency.

    CONCLUSIONS: Scansys and Sirius have small differences and good agreement in the parameters, which can be replaced by each other in clinical practice. Scansys could theoretically be used to extrapolate the implantable contact lens model or could be a new option for anterior segment parameter measurements.

    参考文献
    相似文献
    引证文献
引用本文

赵振波,丁雨溪,唐凯莉,等.两种基于Scheimpflug成像原理的生物测量仪对眼前节参数测量的一致性研究.国际眼科杂志, 2023,23(12):2100-2103.

复制
分享
文章指标
  • 点击次数:
  • 下载次数:
  • HTML阅读次数:
  • 引用次数:
历史
  • 收稿日期:2023-07-05
  • 最后修改日期:2023-11-03
  • 录用日期:
  • 在线发布日期: 2023-11-22
  • 出版日期:
文章二维码